The Model P25S pressure transducer provides high accuracy and reliability at a low cost. The P25S is a variable reluctance transducer with a signal conditioned output of 4 mV/Volt.
Unit of Measure

General Specifications

Absolute Pressure Range

N/A 0 to 1.0 psi through 0 to 1000 psi

Gage Pressure Range

N/A 0 to 2" H2O through 0 to 1000 psi

Differential Pressure Range

N/A 0 to 2" H2O through 0 to 500 psi

Vacuum Pressure Range

N/A 30" Hg through N/A

Overpressure

N/A 20 psi or 200% Full pressure range

Line Pressure (Diff. Units)

N/A 350 psi - Full scale of 5 psi or less. 800 psi - Full scale greater than 5 psi.

Nominal Weight

N/A 25 oz

Sensor Type

N/A Variable Reluctance

Media

N/A Non corrosive liquids or gases in either or both pressure ports

Electrical

Input Voltage

N/A 5 to 10 VDC

Input Current

N/A 350 ohm Full Bridge Std.

Output Voltage

N/A 4 mV/V

Output Noise

N/A 12 microvolt P to P/volt

Output Impedance

N/A 350 ohm resistive

Insulation Resistance

N/A 100 megohms at 100 VDC min.

Adjustment Range

N/A Span: ± 5.0% F.S. min. Zero: ± 5.0% F.S. min.

Performance

Max. Static Error Band

N/A ± 0.3% F.S.

Max. Temperature Error

N/A ± 2.0% F.S.

Max. Regulation

N/A 0.001% F.S. per Volt

Resolution

N/A Infinite

Environmental

Temperature Range

N/A Compensated: -65ºF to 225ºF Operating: 30ºF to 130ºF

Humidity, Rain, Fungus, Sand and Dust

N/A Unaffected

Features

N/A

  • High Accuracy
  • High Stability
  • Input/Output Options
  • Low Power Consumption