The Model P110C variable reluctance absolute or sealed pressure transducer is designed specifically for the petro-chemical industry. All welded construction reduces long term drift and hysteresis and increases accuracy. The pressure cavity is designed with no entrapment areas to facilitate cleaning. This unit features Tavis Corporations highly stable, reliable, accurate 4 to 20 mA electronics.
Unit of Measure

General Specifications

Pressure Range

N/A 0 to 100 through 0 to 10000 psi

Proof Pressure

N/A 200% F.R. to 10000 psi max.

Max. Weight

N/A 3 lb

Sensor Type

N/A Variable Reluctance

Media

N/A Liquids or gases compatible with 17-4 PH SS

External Materials

N/A 17-4 PH and 300 Series SS

Electrical

Input Voltage (At the Transducer)

N/A 17 to 50 VDC

Power Supply Voltage

N/A Transducer Input Voltage + (0.02 x load)

Output Voltage

N/A 4 to 20 mA

Max. Output Noise

N/A 1.0% F.S. peak

Load Resistance

N/A (min. Supply voltage - 17) ˜ 0.02

Insulation Resistance

N/A 100 megohms min. at 100 VDC

Adjustment Range

N/A Span: ± 2% F.S. min. Zero: ± 2% F.S. min.

Performance

Static Error Band

N/A Ranges 5000 psi and up : ± 0.5% F.S. max. Ranges less than 5000 psi : ± 0.2% F.S. max.

Max. Temperature Error

N/A ± 0.75% F.S.

Max. Regulation

N/A 0.025% F.S. per volt

Environmental

Temperature Range

N/A Compensated: 30ºF to +130ºF Operating: -65ºF to +185ºF

Vibration

N/A 35 g s Peak to 2000 Hz

Shock

N/A 100 g s, Peak to 11 ms

Humidity, Rain, Fungus, Sand and Dust

N/A Unaffected

Features

N/A

  • High Accuracy
  • Long Term Stability
  • Cleanable
  • Stable Calibration
  • Vibration Resistant